Home > Products > Gas Control Equipment > PCTRIO Change Over Unit
0 to 100 PSI Pneuchange PCTRIO Base Unit c/w Integrated Battery Alarm
The Pneuchange Change Over unit was developed specifically for use with CO2 and other gas dependent incubators although it is now used in many other applications. Since it''s introduction it has become the unit of choice in thousands of laboratories world-wide. Pneuchange is recommended and supplied by most European incubator manufacturers and distributors and is compatible with all makes of incubator.
Pneuchange operates on a unique pneumatic logic principle requiring no electrical supply except for a battery to operate the change-over alarm. It can be placed anywhere in the gas supply line between the cylinder regulators and the incubator, either free sanding or fixed to a wall using the keyholes on the back of the unit. Inlet pressure gauges indicate the status of the gas cylinders and coloured indicators show which one is active. An outlet pressure regulator and gauge facilitate precise outlet pressure adjustment to the equipment. On depletion of a cylinder the unit automatically feeds from the alternative cylinder with no disruption of supply. Exhausted cylinders can be replaced at leisure. The unit requires no manual reset.
Inlet and outlet ports are 3/8”BSP threaded to accommodate a range of hose-tails to fit the supply tubing. Adaptors for other types of connection are also available.
Description: 0 to 100 PSI Pneuchange Base Unit c/w Integrated Battery Alarm
Extended Description: Two inlet gas cylinder automatic changeover unit providing a continuous supply of gas to a choice of three outputs whilst the exhausted cylinder is replaced. Outlet pressure control is via plug-in OPM (Outlet Pressure Module) regulators (see Other Technical Data). The unit is designed for use with CO2 and other inert gases supplying incubators and similar laboratory equipment. The unit provides a visual indication of active input and an audible indication of changeover
Max. Inlet Pressure: 6.9bar (100psi)
Operating Pressure: 0 to 2.1bar (0 to 30psi) [See also Other Technical Data]
Inlet Connection: 2 off 3/8” BSPP male coned R/H c/w 2 off hosetails to suit 6.4mm (1/4”) ID hose
Outlet Connection: The unit incorporates three self-sealing outlet ports into which special separately supplied outlet pressure modules (OPM’s) can be plugged
Mounting Style: Flush, In-line
Wetted Materials: Brass, nickel-plated brass, steel, polyurethane, technopolymer, Cu-alloy, nylon, nitrile rubber, acetal
Non-wetted Materials: Steel (zinc-plated, chrome-plated), stainless steel, plastic, vinyl, aluminium alloy
Operating Temperature: 0°C to 50°C (32° to 122°F)
Recommended Filtration: 20 micron. User to provide separately [see Other Technical Data].
Weight: 2.700kg (5.95lbs)
Voltage: 9Vdc (battery alarm)
Wattage: 0.072W
CE Mark: Yes
ROHS Compliant: Yes
REACH Compliant: Yes
Clean Room Manufactured: Controlled Environment
Country of Origin: United Kingdom
Any other Technical Data: This unit must only be used for CO2 and other inert gases. The regulator is of the relieving type and so it is recommended that suitable ventilation is provided when the unit is operating. When setting the outlet pressure, always adjust on increasing pressure. Filters are available for incorporation into the Pneuchange inlet lines.